Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 30" X 4" 1-LINE (CS 18) WHITE Model/Catalog Number…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0003-2027
- Category
- Medical devices
- Recall date
- 2026-08-24
- Enforcement report date
- 2026-10-07
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- ASP Global, LLC.
From the FDA Enforcement Report dated 2026-10-07. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0003-2027.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: Expressaire Product Name: DISPOSABLE TOURNIQUET CUFF, 30" X 4" 1-LINE (CS 18) WHITE Model/Catalog Number: 553322 Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No
- Hazard / reason
- Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
- Quantity
- 5015
- Distribution
- US Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.
- Code information
- Lot Code: Model Number: 553322 UDI-DI: (01) 00885914010532 All Lots produced from 01SEP2016 to 01MAR2023: S0922STY, R0310STY, R0801STY, Q0930STY, P0620STY, O0815STY, N0730STY, M0930STY
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Possible model numbers (as listed; not verified as model numbers)
- 553322