VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006

Agency
U.S. Food and Drug Administration
Recall number
Z-0070-2027
Category
Medical devices
Recall date
2026-08-31
Enforcement report date
2026-10-14
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Steris Corporation

From the FDA Enforcement Report dated 2026-10-14. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006
Hazard / reason
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
Quantity
266
Distribution
Worldwide - US Nationwide distribution.
Code information
UDI: 10724995220324 Lot numbers: 260810 - expiration date 08/10/2026 260810A - expiration date 08/10/2026 261106 - expiration date 11/06/2026 270116 - expiration date 01/16/2027
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 10724995220324

Brands