Alphenix INFX-8000C/B5

Agency
U.S. Food and Drug Administration
Recall number
Z-0108-2027
Category
Medical devices
Recall date
2026-08-26
Enforcement report date
2026-10-14
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Canon Medical System, USA, INC.

From the FDA Enforcement Report dated 2026-10-14. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Loosening of the bolts securing the longitudinal movement mechanism may cause the tabletop to shift, resulting in [positional/imaging/patient] misalignment and a longitudinal motion error
Quantity
1 unit
Distribution
US Nationwide distribution in the state of ID.
Code information
Serial no. B5F2532011/UDI: 4987670100192
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 4987670100192

Brands