TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;

Agency
U.S. Food and Drug Administration
Recall number
Z-0117-2027
Category
Medical devices
Recall date
2026-08-28
Enforcement report date
2026-10-14
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
LeMaitre Vascular, Inc.

From the FDA Enforcement Report dated 2026-10-14. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;
Hazard / reason
Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.
Quantity
54
Distribution
US Nationwide distribution in the states of CA, FL, GA, KY, LA, NJ, OH, PA, SC.
Code information
Catalog Number: E1651-88; UDI-DI: 00840663100866; Lot Number: XOT00002181;
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 840663100866

Brands