TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0117-2027
- Category
- Medical devices
- Recall date
- 2026-08-28
- Enforcement report date
- 2026-10-14
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- LeMaitre Vascular, Inc.
From the FDA Enforcement Report dated 2026-10-14. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0117-2027.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- TufTex¿ Over-the-Wire Embolectomy Catheter; Catalog Number: E1651-88;
- Hazard / reason
- Devices failed in process testing related to the amount of pull force it can sustain prior to the balloon components failing.
- Quantity
- 54
- Distribution
- US Nationwide distribution in the states of CA, FL, GA, KY, LA, NJ, OH, PA, SC.
- Code information
- Catalog Number: E1651-88; UDI-DI: 00840663100866; Lot Number: XOT00002181;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 840663100866