RAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML020115

Agency
U.S. Food and Drug Administration
Recall number
Z-0154-2027
Category
Medical devices
Recall date
2026-09-21
Enforcement report date
2026-10-21
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Steris Corporation

From the FDA Enforcement Report dated 2026-10-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
If the vented cap becomes blocked, gas may accumulate and the bottle may swell due to increased internal pressure. In some cases, excess vapor or liquid chemistry may escape from the bottle when the cap is twisted during opening, which could result in user exposure if required personal protective equipment (PPE) is not properly used. Such exposure may result in injury to the user or others in close proximity.
Quantity
374 units
Distribution
Worldwide - US Nationwide distribution.
Code information
UDI: 00724995245658 Lot numbers with expiration dates: 687046 10/5/2027 688017 10/19/2027
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 724995245658

Brands