Medline surgical convenience kits labeled as: VEIN HARVESTING PACK, Medline product number/SKU DYNJ51845D
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0172-2027
- Category
- Medical devices
- Recall date
- 2026-09-02
- Enforcement report date
- 2026-10-21
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2026-10-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0172-2027.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Medline surgical convenience kits labeled as: VEIN HARVESTING PACK, Medline product number/SKU DYNJ51845D
- Hazard / reason
- Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
- Quantity
- 56 kits
- Distribution
- US Nationwide distribution.
- Code information
- UDI/DI 10195327005412 (each) 40195327005413 (case), Lot number 23JMA378
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10195327005412