Medtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly

Agency
U.S. Food and Drug Administration
Recall number
Z-0197-2027
Category
Medical devices
Recall date
2026-08-31
Enforcement report date
2026-10-28
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medtronic Navigation, Inc.

From the FDA Enforcement Report dated 2026-10-28. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Medtronic Biopsy Needle Kit, REF: 9733068, Passive, Sterile EO, Rxonly
Hazard / reason
Sterile Passive spheres and biopsy needle kits contain variation in the reflective material which may affect the ability of the navigation system to detect and track the spheres.
Quantity
3746 units
Distribution
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: Canary Islands, Iceland, Northern Ireland, Martinique, China , Taiwan, Hong Kong,
Code information
UDI: 00643169702943/ Lots: 066500526 066501226 066503326 066504726 066506326 066509826; UDI:00643169030121/Lots: 066500526 066506326
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 643169702943
  • 643169030121

Brands