Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0200-2027
- Category
- Medical devices
- Recall date
- 2026-07-29
- Enforcement report date
- 2026-10-28
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Greiner Bio-One North America, Inc.
From the FDA Enforcement Report dated 2026-10-28. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0200-2027.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
- Hazard / reason
- Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
- Quantity
- 860450
- Distribution
- U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
- Code information
- Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II