Techno-Path Multichem AE Unassayed Level 3, REF:AE623X

Agency
U.S. Food and Drug Administration
Recall number
Z-0207-2027
Category
Medical devices
Recall date
2026-08-14
Enforcement report date
2026-10-28
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
TECHNO-PATH MANUFACTURING LTD.

From the FDA Enforcement Report dated 2026-10-28. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Quantity
832 units
Distribution
US Nationwide distribution in the state of Massachusetts.
Code information
UDI:(01)05391523442824/-Lot # 65705263
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II