Techno-Path Multichem AE Unassayed Level 3, REF:AE623X
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0207-2027
- Category
- Medical devices
- Recall date
- 2026-08-14
- Enforcement report date
- 2026-10-28
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- TECHNO-PATH MANUFACTURING LTD.
From the FDA Enforcement Report dated 2026-10-28. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0207-2027.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
- Quantity
- 832 units
- Distribution
- US Nationwide distribution in the state of Massachusetts.
- Code information
- UDI:(01)05391523442824/-Lot # 65705263
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II