Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Model/Catalog…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0287-2027
- Category
- Medical devices
- Recall date
- 2026-10-05
- Enforcement report date
- 2026-11-11
- Classification
- Class I
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Edwards Lifesciences, LLC
From the FDA Enforcement Report dated 2026-11-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0287-2027.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Generic Name: Edwards Commander M Delivery System Generic Name: Edwards Commander M Delivery System Model/Catalog Number: 9880CM29A Software Version: Not applicable. Product Description: Prosthesis, mitral valve, percutaneously delivered Component: No
- Hazard / reason
- Due to a manufacturing issue which may result in cardiovascular balloon rupturing.
- Quantity
- 164
- Distribution
- U.S.: AZ, CA, CO, FL, GA, IL, KS, LA, MA, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA and WA. O.U.S.: N/A
- Code information
- Lot Code: Model No.: 9880CM29A; UDI-DI: 00690103219777; Recalled Lot Number(s): 67718561, 67804014, 67887043, 67887044.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class I
Identifiers
- Barcodes (UPC/GTIN)
- 690103219777
- Model numbers
- 9880CM29A