Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103

Agency
U.S. Food and Drug Administration
Recall number
Z-0317-2027
Category
Medical devices
Recall date
2026-09-14
Enforcement report date
2026-11-11
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Accriva Diagnostics, Inc.

From the FDA Enforcement Report dated 2026-11-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.
Quantity
28710
Distribution
Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.
Code information
UDI-DI: 10711234103012, Lot(Expiration): B6JCA005(02/28/2027), C6JCA008(03/31/2027), D6JCA011(04/30/2027), K5JCA041(10/31/2026), C6JCA007(03/31/2027)
Recall type
Voluntary: Firm initiated
Product type
Devices