Embletta MPR PG Multi Parameter Recorder Sleep Monitor, Model/Catalog Number: 2003011

Agency
U.S. Food and Drug Administration
Recall number
Z-0319-2027
Category
Medical devices
Recall date
2026-09-29
Enforcement report date
2026-11-11
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
NATUS MEDICAL INCORPORATED

From the FDA Enforcement Report dated 2026-11-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
The device may be unable to correctly format its internal memory. When the internal memory cannot be correctly formatted, previously recorded studies may remain on the device and may be updated with incorrect patient information from a subsequent study. The issue may also prevent studies from initiating recording successfully or may cause recordings to end prematurely, resulting in incomplete or unavailable diagnostic data and the potential need to repeat a patient study.
Quantity
288 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, China, Italy, Portugal.
Code information
UDI-DI: 00382830023764; Serial Numbers: EBMPG-C1013678 EBMPG-C1013679 EBMPG-C1013682 EBMPG-C1013690 EBMPG-C1013691 EBMPG-C1013697 EBMPG-C1013700 EBMPG-C1013701 EBMPG-C1013703 EBMPG-C1013704 EBMPG-C1013707 EBMPG-C1013708 EBMPG-C1013748 EBMPG-C1013778 EBMPG-C1013779 EBMPG-C1013876 EBMPG-C1013566 EBMPG-C1013620 EBMPG-C1013632 EBMPG-C1013594 EBMPG-C1013596 EBMPG-C1013597 EBMPG-C1013599 EBMPG-C1013602 EBMPG-C1013614 EBMPG-C1013615 EBMPG-C1013617 EBMPG-C1013618 EBMPG-C1013414 EBMPG-C1013415 EBMPG-C1013416 EBMPG-C1013417 EBMPG-C1013448 EBMPG-C1013418 EBMPG-C1013591 EBMPG-C1013621 EBMPG-C1013629 EBMPG-C1013634 EBMPG-C1013636 EBMPG-C1013637 EBMPG-C1013643 EBMPG-C1013648 EBMPG-C1013655 EBMPG-C1013665 EBMPG-C1013609 EBMPG-C1013650 EBMPG-C1013652 EBMPG-C1013666 EBMPG-C1013670 EBMPG-C1013649 EBMPG-C1013293 EBMPG-C1013376 EBMPG-C1013387 EBMPG-C1013392 EBMPG-C1013398 EBMPG-C1013616 EBMPG-C1013633 EBMPG-C1013630 EBMPG-C1013638 EBMPG-C1013627 EBMPG-C1013541 EBMPG-C1013661 EBMPG-C1013667 EBMPG-C1013495 EBMPG-C1013572 EBMPG-C1013577 EBMPG-C1013578 EBMPG-C1013595 EBMPG-C1013610 EBMPG-C1013642 EBMPG-C1013653 EBMPG-C1013723 EBMPG-C1013734 EBMPG-C1013587 EBMPG-C1013590 EBMPG-C1013606 EBMPG-C1013611 EBMPG-C1013613 EBMPG-C1013641 EBMPG-C1013676 EBMPG-C1013542 EBMPG-C1013574 EBMPG-C1013583 EBMPG-C1013584 EBMPG-C1013672 EBMPG-C1013673 EBMPG-C1013680 EBMPG-C1013681 EBMPG-C1013683 EBMPG-C1013685 EBMPG-C1013686 EBMPG-C1013687 EBMPG-C1013688 EBMPG-C1013689 EBMPG-C1013693 EBMPG-C1013694 EBMPG-C1013695…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 382830023764

Brands