Mobile column TruSystem 7500 U, Product Code 1730720
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0333-2027
- Category
- Medical devices
- Recall date
- 2026-09-14
- Enforcement report date
- 2026-11-11
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Baxter Healthcare Corporation
From the FDA Enforcement Report dated 2026-11-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0333-2027.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
- Quantity
- 1 unit
- Distribution
- US Nationwide distribution in the states of IL, MO, TX.
- Code information
- UDI-DI: 00887761974388; Serial number: 110183858
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 887761974388