Mobile column TruSystem 7500 U, Product Code 1730720

Agency
U.S. Food and Drug Administration
Recall number
Z-0333-2027
Category
Medical devices
Recall date
2026-09-14
Enforcement report date
2026-11-11
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Baxter Healthcare Corporation

From the FDA Enforcement Report dated 2026-11-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
The main lift may unexpectedly drop from its set position to its lowest position during use. The tabletop may not be able to be raised again after the drop occurs. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others.
Quantity
1 unit
Distribution
US Nationwide distribution in the states of IL, MO, TX.
Code information
UDI-DI: 00887761974388; Serial number: 110183858
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 887761974388

Brands