Unity Knee ASC Modular Handle included in the Corin UNITY ASC INSTRUMENT SYSTEM. Part Number: 1232040.

Agency
U.S. Food and Drug Administration
Recall number
Z-0356-2027
Category
Medical devices
Recall date
2026-10-02
Enforcement report date
2026-11-11
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
CORIN MEDICAL, LTD.

From the FDA Enforcement Report dated 2026-11-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Potential malfunction of the Modular Handles where the locking mechanism may fail. Any mating instruments already engaged to the handle will not be able to be disengaged, and the handle will be unable to fitted securely to any unengaged mating instruments.
Quantity
22 units
Distribution
Worldwide - US Nationwide distribution in the states of CA, GA, IL, MI, NC, RI, TX, UT and the countries of Belgium, France, Germany, Hungary.
Code information
Part Number: 1232040. UDI-DI/GTIN; 05056139239656. Lot Code: 097921.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands