Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-0571-2026
Category
Medical devices
Recall date
2025-08-19
Enforcement report date
2025-11-26
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Greiner Bio-One North America, Inc.

From the FDA Enforcement Report dated 2025-11-26. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product Description: Greiner VACUETTE Blood Collection Tube with Clot Activator and Gel Separator and Standard Stopper (16 x 100 mm, 8ml)
Hazard / reason
Some of the Vacuette¿ CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat testing. This may result in delayed diagnosis, cause a delay in clinical decision-making or treatment and in some rare cases, yield incorrect results from less stable parameters.
Quantity
1,980,000
Distribution
U.S.: AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, VA and WY. O.U.S.: None
Code information
Lot Code: Lot/Serial Numbers: Item# 455071P Lot# B250437N UDI-DI code: 29120017573061
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 29120017573061

Brands