Quantum Intersegmental Table, Model/Catalog Number: Q400; heated massage table

Agency
U.S. Food and Drug Administration
Recall number
Z-0587-2026
Category
Medical devices
Recall date
2025-10-09
Enforcement report date
2025-12-03
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Pivotal Health Solutions, Inc.

From the FDA Enforcement Report dated 2025-12-03. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
Quantity
34 units
Distribution
Distribution to US nationwide, Canada and Japan.
Code information
All serial numbers.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands