Quantum Intersegmental Table, Model/Catalog Number: AMQ400H; heated massage table
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0588-2026
- Category
- Medical devices
- Recall date
- 2025-10-09
- Enforcement report date
- 2025-12-03
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Pivotal Health Solutions, Inc.
From the FDA Enforcement Report dated 2025-12-03. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0588-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- The heat function may become damaged over time due to patients pressing on the areas with heat function components. Damage to components can lead to overheating which can lead to smoldering and melting of the vinyl cushion.
- Quantity
- 464 units
- Distribution
- Distribution to US nationwide, Canada and Japan.
- Code information
- All serial numbers.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II