FlexLab (FLX); Version: FLX-217-10;

Agency
U.S. Food and Drug Administration
Recall number
Z-0595-2026
Category
Medical devices
Recall date
2025-10-30
Enforcement report date
2025-12-03
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Inpeco S.A.

From the FDA Enforcement Report dated 2025-12-03. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
FlexLab (FLX); Version: FLX-217-10;
Hazard / reason
The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.
Quantity
1 unit (OUS only)
Distribution
US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.
Code information
UDI-DI: 07640172340004 Serial number: FLX.0226;
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 7640172340004

Brands