FlexLab (FLX); Version: FLX-217-40;
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0596-2026
- Category
- Medical devices
- Recall date
- 2025-10-30
- Enforcement report date
- 2025-12-03
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Inpeco S.A.
From the FDA Enforcement Report dated 2025-12-03. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0596-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- FlexLab (FLX); Version: FLX-217-40;
- Hazard / reason
- The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
- Quantity
- 14 units (1 US, 13 OUS)
- Distribution
- US Domestic: MI. OUS International: Australia, Finland, France, Germany, Italy, Portugal, Spain, and UK.
- Code information
- UDI-DI: 07640172344002; Serial number: FXX.0004, FXX.0006, FXX.0009, FXX.0022, FXX.0023, FXX.0024, FXX.0027, FXX.0028, FXX.0034, FXX.0036, FXX.0041, FXX.0042, FXX.0043, FXX.0046;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 7640172344002