Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0597-2026
- Category
- Medical devices
- Recall date
- 2025-10-27
- Enforcement report date
- 2025-12-17
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Philips North America
From the FDA Enforcement Report dated 2025-12-17. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0597-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
- Hazard / reason
- Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
- Quantity
- 110
- Distribution
- U.S.
- Code information
- Model Number 722234 Commercial Name: Azurion 7 M20 How to identify: Philips Reference C&R 2025-IGT-BST-016
- Recall type
- FDA Mandated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Possible model numbers (as listed; not verified as model numbers)
- 722234