Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM

Agency
U.S. Food and Drug Administration
Recall number
Z-0597-2026
Category
Medical devices
Recall date
2025-10-27
Enforcement report date
2025-12-17
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Philips North America

From the FDA Enforcement Report dated 2025-12-17. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Hazard / reason
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
Quantity
110
Distribution
U.S.
Code information
Model Number 722234 Commercial Name: Azurion 7 M20 How to identify: Philips Reference C&R 2025-IGT-BST-016
Recall type
FDA Mandated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Possible model numbers (as listed; not verified as model numbers)
  • 722234

Brands