BioFire Respiratory Panel 2.1, IVD, REF: 423742 (30 test kit)

Agency
U.S. Food and Drug Administration
Recall number
Z-0602-2026
Category
Medical devices
Recall date
2025-10-22
Enforcement report date
2025-12-03
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
BioFire Diagnostics, LLC

From the FDA Enforcement Report dated 2025-12-03. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Due to unintended movement of pouch chemistry, nucleic acid test may result in elevated risk of false negative results and risk of control failures.
Quantity
97 Kits (2,910 pouches)
Distribution
US:AZ, FL,GA,ID NC, NH, NJ, NY, OR, PA, SC OUS: None
Code information
UDI: 00815381020529 /Lot # 2649724 Expanded Lot#: 1471725
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 815381020529

Brands