Serotonin Test

Agency
U.S. Food and Drug Administration
Recall number
Z-0815-2026
Category
Medical devices
Recall date
2025-11-03
Enforcement report date
2025-12-10
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
GET TESTED INTERNATIONAL AB

From the FDA Enforcement Report dated 2025-12-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Distribution without premarket approval/clearance.
Quantity
12 units
Distribution
US Nationwide distribution.
Code information
EAN: 616612785718; SKU: 1SE; UDI-DI: None; Lot/Serial Number: All Lots;
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 616612785718

Brands