Phantom Fibula Nail, 2.8mm x 130mm, Right, Ti, Fibula Nail, REF: P36-128-130R-S
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0871-2026
- Category
- Medical devices
- Recall date
- 2025-10-13
- Enforcement report date
- 2025-12-10
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Paragon 28, Inc.
From the FDA Enforcement Report dated 2025-12-10. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0871-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Fibula nail system intended for use in fixation of fibular fractures and osteotomies, has out of specification thread depths on the fibula nail, so implant may not be able to mate properly with the associated mounting bolt. If nail (implant) cannot properly mate with instrumentation, the surgery potentially will need to be aborted.
- Quantity
- 20
- Distribution
- US Nationwide distribution in the states of NJ, WA, NY, MD, CA, FL, NM, TX, IL, OH, NV, TN, LA.
- Code information
- UDI-DI: 00889795133670. Lot: 501182207125. Expiration: March 13th, 2030
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 889795133670