Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

Agency
U.S. Food and Drug Administration
Recall number
Z-0887-2026
Category
Medical devices
Recall date
2025-11-06
Enforcement report date
2025-12-17
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Bard Peripheral Vascular Inc

From the FDA Enforcement Report dated 2025-12-17. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.
Quantity
1240
Distribution
US Nationwide distribution in the states of CA, MN, IN, MI, LA.
Code information
UDI/Lot/Expiration: (01)10885403108365(17)271130(10)0001623320/0001623320/20271130; (01)10885403108365(17)271130 (10)0001623069/0001623069/20271130
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II