Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0890-2026
- Category
- Medical devices
- Recall date
- 2025-11-10
- Enforcement report date
- 2025-12-17
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
From the FDA Enforcement Report dated 2025-12-17. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0890-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Endo Model Knee Fusion Nail SK, Knee Fusion Nail, Coupling Component, right, modular, L= 68 mm, Taper 10/12. Knee component of limb salvage replacement devices.
- Hazard / reason
- The device has been delivered with a preassembled locking screw that had been inserted from the wrong side into the male taper of the femoral component.
- Quantity
- 18 units
- Distribution
- US distribution to states of: AL, GA, and TX.
- Code information
- Item Number: 15-0028/08; UDI-DI: 04026575175222; Lot Number: 2535004;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 4026575175222