SCISSORS, OR, SHARP/BLUNT, 5.5", STERILE, Item Number DYND04000
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0921-2026
- Category
- Medical devices
- Recall date
- 2025-11-13
- Enforcement report date
- 2025-12-24
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medline Industries, LP
From the FDA Enforcement Report dated 2025-12-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0921-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- The items may have weak packaging seals that may result in a breach of sterility if the seal fails and may lead to an increased risk of infection if product is utilized.
- Quantity
- 56 units
- Distribution
- US Nationwide distribution in the states CA, CO, IL, KS, and MI.
- Code information
- DYND04000, UDI-DI: 10193489191349(each), 20193489191346(case), Lot Number: 2023102590
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10193489191349