Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Agency
U.S. Food and Drug Administration
Recall number
Z-0923-2026
Category
Medical devices
Recall date
2025-11-18
Enforcement report date
2025-12-31
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
AGFA Healthcare Corp.

From the FDA Enforcement Report dated 2025-12-31. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800
Hazard / reason
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
Quantity
35
Distribution
U.S. Nationwide distribution.
Code information
N/A
Recall type
FDA Mandated
Product type
Devices
Classification (FDA wording)
Class II