HemosIL SynthAFax. Partial Thromboplastin Time Tests.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0938-2026
- Category
- Medical devices
- Recall date
- 2025-11-18
- Enforcement report date
- 2025-12-24
- Classification
- Class III
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Instrumentation Laboratory
From the FDA Enforcement Report dated 2025-12-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0938-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Recalled lots were manufactured with double the amount of preservative concentration.
- Quantity
- 4,506 units
- Distribution
- Domestic: CA, CO, FL, NY; Foreign: AR, AT, CA, CZ, DE, DK, ES, FR, GB, IE, IL, IN, IT, JP, LK, NL, PL, PT, RO, RS, TH, VN.
- Code information
- Part No. 0020007400; UDI: 08426950087649; Lot No. (Exp. Date) N0542231(2026-06-30), N0250088(2027-04-30, N0452249 (2027-06-30).
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class III
Identifiers
- Barcodes (UPC/GTIN)
- 8426950087649