Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54544-18-PK…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0952-2026
- Category
- Medical devices
- Recall date
- 2025-11-07
- Enforcement report date
- 2025-12-24
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Rocket Medical Plc
From the FDA Enforcement Report dated 2025-12-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0952-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter Procedure Tray Model/Catalog Number: R54544-18-PK Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices. Component: N/A
- Hazard / reason
- IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
- Quantity
- 630 units
- Distribution
- Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
- Code information
- Model/Catalog Number: R54544-18-PK; UDI-DI: 050552709TF06DXY; Lot Code: 496899, 500956, 497076, 501169, 497584, 501188, 499227, 502430, 499417, 502951, 499461, 503306, 499897, 503430, 499898, 503529, 499899, 504180, 499908, 504543, 500407, 504823, 500477;
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II