Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter with Safety Guard Model/Catalog Number: R54544-18-SG…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-0953-2026
Category
Medical devices
Recall date
2025-11-07
Enforcement report date
2025-12-24
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Rocket Medical Plc

From the FDA Enforcement Report dated 2025-12-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Brand Name: Rocket Product Name: Rocket 18Fg Seldinger Catheter with Safety Guard Model/Catalog Number: R54544-18-SG Software Version: N/A Product Description: Chest tube manufactured from PVC, with fenestrations and a barium stripe to the indicated French gauge size. Supplied with a procedure pack to allow percutaneous insertion, supplied with connectors to allow attachment to chest drain bottle devices.
Hazard / reason
IFU and device kit labeling contains misleading statements leading to improper securing of the device resulting in the potential to fall out during use.
Quantity
110 units
Distribution
Domestic: CO, CT, IL, MA, ME, NH, NJ, NM, OH, RI, SC, TX; International: UAE, Austria, Australia, Canada, Germany, Denmark, Finland, France, United Kingdom, Greece, Hong Kong, Ireland, Israel, India, Italy, Jersey, Kenya, Sri Lanka, Malta, Netherlands, Norway, New Zealand, Portugal, Saudi Arabia, Singapore, Slovenia, South Africa
Code information
Model/Catalog Number: R54544-18-SG; UDI-DI: 050552709TF06DXY; Lot Code: 496809, 501616, 497346, 501710, 497665, 502431, 498972, 503735, 499909, 503937, 500782, 504573, 501256, 504824;
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands