InPen App, Model/CFN Number: MMT-8061 (Android Users)

Agency
U.S. Food and Drug Administration
Recall number
Z-0958-2026
Category
Medical devices
Recall date
2025-11-13
Enforcement report date
2025-12-24
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Medtronic MiniMed, Inc.

From the FDA Enforcement Report dated 2025-12-24. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
When app is uninstalled and reinstalled, insulin pen software issue causes Choose Notification Style Screen to not show during setup so users can't choose to allow notifications to override phone settings when on mute/Do Not Disturb, and previously set up override permission is deleted, so audible and vibratory notifications not received, which may lead to delayed insulin therapy, hyperglycemia.
Quantity
8251
Distribution
US: IL, AZ, MN, NE, KS, TN, CA, FL, TX, WA, MO, NC, NY, WI, IN, ME, CT, RI, MD, PA, OK, KY, MA, DE, UT, ID, VA, MI, OH, LA, NJ, IA, MT, HI, AR, OR, SC, AL, CO, WY, VT, NV, GA, MS, SD, AK, WV, NH, ND, NM. OUS: Austria, Belgium, Chile, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Hungary, Iceland, Ireland, Israel, Italy, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, South Africa, Spain, Sweden, Switzerland, United Kingdom
Code information
Software Versions: 7.5.0, 7.5.1, and 8.0.0. UDI-DI: 00763000974596
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 763000974596

Brands