LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0965-2026
- Category
- Medical devices
- Recall date
- 2025-11-24
- Enforcement report date
- 2025-12-31
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- ICU Medical, Inc.
From the FDA Enforcement Report dated 2025-12-31. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0965-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02
- Hazard / reason
- For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.
- Quantity
- 7 customers
- Distribution
- US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.
- Code information
- version 2.2.1 UDI-DI 00887709127227 Issue affects customers who migrated from LifeShield DLM v2.1x to LifeShield DLM v2.2.1 and have any version of a Plum Duo v1.1 drug library that was approved or locked for review before the migration.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 887709127227