LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0966-2026
- Category
- Medical devices
- Recall date
- 2025-11-24
- Enforcement report date
- 2025-12-31
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- ICU Medical, Inc.
From the FDA Enforcement Report dated 2025-12-31. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0966-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
- Hazard / reason
- Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.
- Quantity
- 12 customers
- Distribution
- US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.
- Code information
- LifeShield Infusion Safety Software Suite v2.2 UDI-DI 00887709127227
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 887709127227