Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0996-2026
- Category
- Medical devices
- Recall date
- 2025-11-20
- Enforcement report date
- 2025-12-31
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Medtronic Neuromodulation
From the FDA Enforcement Report dated 2025-12-31. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0996-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Restore Clinician Programmer Application (CP App), Model A71100 used with the following programmers: CT900A PROG CT900A CLINICIAN TABLET US CT900B PROG CT900B CLINICIAN TABLET EU -UK CT900C PROG CT900C CLINICIAN TABLET UK CT900D PROG CT900D CLINICIAN TABLET GLOBAL CT900E PROG CT900E CLINICIAN TABLET GLBL SMSNG CT900F PROG CT900F CLINICIAN TABLET GLBL SMSNG
- Hazard / reason
- Software issue where a Device Reset message displayed on the app was unable to be cleared. In rare cases, this resulted in an inability to resume therapy, and the patient experienced a recurrence of underlying pain symptoms. Resolution would require surgical replacement of the INS.
- Quantity
- 7667 devices
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Aruba, Australia, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Finland, France, French Polynesia, Germany, Greece, Guatemala, Hong Kong, Hungary, Iceland, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Malta, Mauritius, Mexico, Namibia, Netherlands, New Zealand, Norway, Oman, Panama, Peru, Poland, Portugal, Qatar, Reunion, Russia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Uruguay.
- Code information
- CP App: UDI-DI 00763000273668 all software versions prior to v.1.0.4489 are affected The implanted devices are not affected.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 763000273668
- Model numbers
- A71100