LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-0999-2026
- Category
- Medical devices
- Recall date
- 2025-12-05
- Enforcement report date
- 2026-01-14
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Siemens Medical Solutions USA, Inc
From the FDA Enforcement Report dated 2026-01-14. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-0999-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the collimation may be incorrect.
- Quantity
- U.S.: 1, OUS: 95
- Distribution
- U.S. and OUS
- Code information
- (01)04056869269931(21)10330
- Recall type
- FDA Mandated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Possible model numbers (as listed; not verified as model numbers)
- 11574100
- 10762471