BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1008-2026
- Category
- Medical devices
- Recall date
- 2025-12-05
- Enforcement report date
- 2026-01-07
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Beaver-Visitec International, Inc.
From the FDA Enforcement Report dated 2026-01-07. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1008-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Hazard / reason
- Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
- Quantity
- 5 units
- Distribution
- US Nationwide distribution in the states of FL, MI, OK, TX, WI.
- Code information
- Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Model numbers
- OME6000U