BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

Agency
U.S. Food and Drug Administration
Recall number
Z-1008-2026
Category
Medical devices
Recall date
2025-12-05
Enforcement report date
2026-01-07
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Beaver-Visitec International, Inc.

From the FDA Enforcement Report dated 2026-01-07. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
Quantity
5 units
Distribution
US Nationwide distribution in the states of FL, MI, OK, TX, WI.
Code information
Model Number: OME6000U. UDI: (01)00886158037576(11)250513(21). Serial Numbers: 77UBV500003, 77UBV500004, 77UBV500005, 77UBV500006, 77UBV500010.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Model numbers
  • OME6000U

Brands