MEDLINE beds labeled as: Full Electric Low Lightweight Homecare Bed, REF MDR107003LO

Agency
U.S. Food and Drug Administration
Recall number
Z-1016-2026
Category
Medical devices
Recall date
2025-11-26
Enforcement report date
2026-01-28
Classification
Class I
Status as reported by FDA
Ongoing
Manufacturer / firm
Medline Industries, LP

From the FDA Enforcement Report dated 2026-01-28. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Hazard / reason
Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
Quantity
15581 units
Distribution
US, CA, PA, GUAM, VI
Code information
UDI/DI 40080196325901, All lots
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class I

Identifiers

Barcodes (UPC/GTIN)
  • 40080196325901

Brands