VANTAGE GALAN 3T Model MRT-3020/MEXL-3020

Agency
U.S. Food and Drug Administration
Recall number
Z-1017-2026
Category
Medical devices
Recall date
2025-12-19
Enforcement report date
2026-01-21
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Canon Medical System, USA, INC.

From the FDA Enforcement Report dated 2026-01-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
Hazard / reason
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
Quantity
34 systems
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
Code information
Model Number: MRT-3020/MEXL-3020 UDI-DI code: 04987670102615 CMSC Model Number/Model Number/Serial Numbers/SID/UDI code: MRT-3020/4A MEXL-3020/5D 5DB2152002 30058153 (01)04987670102615(21)4AA2152001 MRT-3020/4A MEXL-3020/5D 5DC2262005 30075456 (01)04987670102615(21)4AB2262002 MRT-3020/4A MEXL-3020/5D 5DE2452007 30097122 (01)04987670102615(21)4AC2452003 MRT-3020/5A MEXL-3020/5D 5DB2172003 30047848 (01)04987670102615(21)5AA2172001 MRT-3020/6A MEXL-3020/7D 7DE2482014 30102813 (01)04987670102615(21)6AA2482002 MRT-3020/6A MEXL-3020/7D 7DE2472012 30108012 (01)04987670102615(21)6AA2472001 MRT-3020/7A MEXL-3020/7D 7DD2392008 30100563 (01)04987670102615(21)7AA2392001 MRT-3020/A4 MEXL-3020/D5 D5B2152013 30060527 (01)04987670102615(21)A4D2152004 MRT-3020/A4 MEXL-3020/D5 D5B2212019 30064817 (01)04987670102615(21)A4D2212006 MRT-3020/A4 MEXL-3020/D5 D5A2082005 30069306 (01)04987670102615(21)A4C2082002 MRT-3020/A4 MEXL-3020/D5 D5A2092007 30071480 (01)04987670102615(21)A4C2092003 MRT-3020/A4 MEXL-3020/D5 D5E24X2043 30072643 (01)04987670102615(21)A4F24X2010 MRT-3020/A4 MEXL-3020/D5 D5B2192016 30085151 (01)04987670102615(21)A4D2192005 MRT-3020/A4 MEXL-3020/D5 D5C2262022 30098384 (01)04987670102615(21)A4E2262008 MRT-3020/A4 MEXL-3020/D5 D5C2262021 30100803 (01)04987670102615(21)A4E2262007 MRT-3020/A4 MEXL-3020/D5 D5E2492042 30109383 (01)04987670102615(21)A4F2492009 MRT-3020/A5 MEXL-3020/S5 S5B18X2046 30007763 (01)04987670102615(21)A5B18X2011 MRT-3020/A5 MEXL-3020/S5 S5A17X2033 30031740…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 4987670102615
Model numbers
  • MRT-3020/MEXL-3020

Brands