MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1024-2026
- Category
- Medical devices
- Recall date
- 2025-12-10
- Enforcement report date
- 2026-01-21
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- REPLIGEN CORPORATION
From the FDA Enforcement Report dated 2026-01-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1024-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
- Hazard / reason
- Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
- Quantity
- 33
- Distribution
- U.S.
- Code information
- PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
- Recall type
- FDA Mandated
- Product type
- Devices
- Classification (FDA wording)
- Class II