MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-1024-2026
Category
Medical devices
Recall date
2025-12-10
Enforcement report date
2026-01-21
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
REPLIGEN CORPORATION

From the FDA Enforcement Report dated 2026-01-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
MAVERICK¿ Measurement Module Spectrometer PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Hazard / reason
Potential Elevated Laser Output A disconnected cable or a poorly crimped cable under mechanical bending stress can cause loss of electrical contact. At one connection on the board, the failure can drive the laser output to approximately 1,600 mW. Potential Unintended Laser Firing A disconnected, or poorly crimped cable can falsely signal a fire command during measurement cycles. At a secondary location on the board, this fault can lead to unexpected laser activation during operation. Safety logic prevents firing when the system is idle.
Quantity
33
Distribution
U.S.
Code information
PN 820-01367 (originally part of a kit PN 820-01519) Finished Goods PN 820-01496
Recall type
FDA Mandated
Product type
Devices
Classification (FDA wording)
Class II

Brands