Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1037-2026
- Category
- Medical devices
- Recall date
- 2025-12-12
- Enforcement report date
- 2026-01-14
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Integra LifeSciences Corp. (NeuroSciences)
From the FDA Enforcement Report dated 2026-01-14. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1037-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Codman CereLink ICP Sensor Basic Kit. Catalog Number: 826850. The Codman CereLink ICP Sensor Basic Kit consists of the CereLink ICP Sensor (ICP Sensor) and a 14-gauge Tuohy needle with stylet. The ICP Sensor is a nylon tube with a microminiature strain gauge pressure transducer (sensing element) mounted at one end and an electrical connector at the other end. It is designed for use with a Codman intracranial pressure monitoring device. The ICP Sensor Basic Kit is designed for use with the Codman Cranial Hand Drill. The drill facilitates access to the intraparenchymal area. The drill is also available as a component of the Codman Cranial Access Kit. Use of the ICP Sensor Basic Kit is indicated when direct ICP monitoring is required. The kit is indicated for use in both subdural and intraparenchymal pressure monitoring applications only.
- Hazard / reason
- Potential corrosion stains on the surface of the 14-gauge Tuohy Needle that is included in the Microsensor and Cerelink ICP Kits.
- Quantity
- 5,116 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX UT, VA WA and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia (Plurinational State of), Brazil, Canada, Chile, China, Colombia, Cyprus, Czechia, France, Georgia, Germany, Greece, Hong Kong, Iceland, India, Ireland, Israel, Italy, Japan, Kenya, Korea (the Republic of), Lithuania, Nepal, New Zealand, Poland, Portugal, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Taiwan (Province of China), Thailand, Turkey, United Arab Emirates (the), United Kingdom of Great Britain and Northern Ireland (the).
- Code information
- Catalog Number: 826850. UDI-DI: 10381780520672. Lot Numbers: 7261958, 7266691, 7266692, 7288060, 7293837, 7293836, 7306040, 7306041, 7306042, 7306043, 7306044, 7306045, 7306046, 7306047, 7306050, 7306051, 7306052, 7307670, 7307671, 7307672, 7325482, 7307675, 7325481, 7325483, 7325484, 7307673, 7307674, 7325485, 7325486, 7336611, 7336612, 7336614, 7336623, 7336613, 7336615, 7336616, 7336618, 7336619, 7336620, 7336621, 7336622, 7306048, 7306049, 7336626, 7336627, 7336629, 7347166, 7336628, 7346295, 7346297, 7346299, 7346301, 7365434, 7365435, 7392019, 7396607, 7396608, 7396609, 7396610, 7392034, 7392067, 7405790, 7405791, 7405792, 7405793, 7406633, 7406634, 7406636, 7406637, 7406639, 7406640, 7406641, 7406642, 7406635, 7406638, 7408813, 7408815, 7408818, 7408814, 7408816, 7468726, 7468727, 7411801, 7411802, 7414641, 7414642, 7468731, 7468732, 7468745, 7468746, 7468741, 7468742, 7468743, 7468744, 7468771, 7468769, 7468770, 7468773, 7468774, 7468775, 7482179, 7482180, 7482181, 7482182, 7482183, 7482184, 7482185, 7482186, 7468772, 7482154, 7482155, 7482156, 7482157, 7468801, 7468804, 7468805, 7468806, 7482136, 7468802, 7468803, 7482132, 7482134, 7482135, 7482137, 7482159, 7482158, 7482160, 7482161, 7495593, 7495594, 7495596, 7495597, 7495598, 7495599, 7482212, 7495600, 7495601, 7495602, 7495603, 7495604, 7495605, 7495606, 7495607, 7495608, 7495635, 7495636, 7495637, 7495638, 7495639, 7495640, 7495641, 7495642, 7495643, 7495666, 7495644, 7495664, 7482213, 7495667, 7503597,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 10381780520672