MAXXUS, System, Tomography, Computed, Emission
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1047-2026
- Category
- Medical devices
- Recall date
- 2025-12-10
- Enforcement report date
- 2026-01-21
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
From the FDA Enforcement Report dated 2026-01-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1047-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- MAXXUS, System, Tomography, Computed, Emission
- Hazard / reason
- GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service (herein referred to as Impacted Systems and listed in the Affected Product Details section of this letter) could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury. There have been no reports of detector falls or injuries as a result of this potential issue. The Impacted Systems affected by this notice are those that have not been manufactured for several years. GE HealthCare has previously communicated End of Guaranteed Service for these Impacted Systems.
- Quantity
- 38 units
- Distribution
- US, Brazil, Egypt, Canada, Mexico, Turkey, Italy, Sweden, Switzerland, France, Spain Chile, Germany, Iran, Greece
- Code information
- UDI/DI Not applicable: All serial numbers in distribution
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II