Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1079-2026
- Category
- Medical devices
- Recall date
- 2025-12-08
- Enforcement report date
- 2026-01-21
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
From the FDA Enforcement Report dated 2026-01-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1079-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Philips Azurion Systems R3.0. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Number: 722229. 2. Azurion 3 M15. Model Number: 722230. 3. Azurion 5 M12. Model Number: 722231. 4. Azurion 5 M20. Model Number: 722232. 5. Azurion 7 B12. Model Number: 722235. 6. Azurion 7 B20. Model Number: 722236. 7. Azurion 7 M12. Model Number: 722233. 8. Azurion 7 M20. Model Number: 722234.
- Hazard / reason
- Nine (9) identified software issues which may result in the following: loss of X-ray after Pedal Tap, loss of X-ray after Phase Fault, missing live X-ray on the FlexVision monitor, system continues to restart, AMD triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after start-up, misalignment of Marker Tool Overlay, longitudinal position error.
- Quantity
- 990 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Chile, Colombia, Cote de'Ivore, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Germany, Greece, Honduras, Hong Kong, India, Indonesia, Iraq, Israel, Italy, Japan, Kazakhstan, Kosovo, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Mauritius, Morocco, Nepal, Netherlands, New Zealand, Nicaragua, Norway, Oman, Palestine, Philippines, Poland, Portugal, R¿union, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkiye, Ukraine, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam, Yemen.
- Code information
- Software Version: R3.0. 1. Model Number: 722229. UDI Numbers: (01)00884838116726(21)22, (01)00884838116726(21)15, (01)00884838116726(21)16, (01)00884838116726(21)6, (01)00884838116726(21)8, (01)00884838116726(21)4, (01)00884838116726(21)24, (01)00884838116726(21)26, (01)00884838116726(21)20, (01)00884838116726(21)25, (01)00884838116726(21)11, (01)00884838116726(21)21, (01)00884838116726(21)13, (01)00884838116726(21)7, (01)00884838116726(21)10, (01)00884838116726(21)2, (01)00884838116726(21)3, (01)00884838116726(21)29. 2. Model Number: 722230. UDI Numbers: (01)00884838116733(21)2, (01)00884838116733(21)40, (01)00884838116733(21)9, (01)00884838116733(21)1, (01)00884838116733(21)38, (01)00884838116733(21)10, (01)00884838116733(21)45, (01)00884838116733(21)7, (01)00884838116733(21)27, (01)00884838116733(21)28, (01)00884838116733(21)39, (01)00884838116733(21)4, (01)00884838116733(21)5, (01)00884838116733(21)23, (01)00884838116733(21)41, (01)00884838116733(21)20, (01)00884838116733(21)19, (01)00884838116733(21)13, (01)00884838116733(21)15, (01)00884838116733(21)12, (01)00884838116733(21)22. 3. Model Number: 722231. UDI Numbers: (01)00884838116740(21)13, (01)00884838116740(21)29, (01)00884838116740(21)50, (01)00884838116740(21)21, (01)00884838116740(21)19, (01)00884838116740(21)12, (01)00884838116740(21)25, (01)00884838116740(21)4, (01)00884838116740(21)1, (01)00884838116740(21)5, (01)00884838116740(21)2, (01)00884838116740(21)73, (01)00884838116740(21)72,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Possible model numbers (as listed; not verified as model numbers)
- 722229
- 722230
- 722231
- 722232
- 722235
- 722236
- 722233
- 722234