Brand Name: PALACOS Product Name: PALACOS R pro 40; PALACOS R pro 80 Model/Catalog Number: 5081286; 5081287 Software…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1080-2026
- Category
- Medical devices
- Recall date
- 2025-12-09
- Enforcement report date
- 2026-01-21
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Heraeus Medical GmbH (Dental Division)
From the FDA Enforcement Report dated 2026-01-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1080-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: PALACOS Product Name: PALACOS R pro 40; PALACOS R pro 80 Model/Catalog Number: 5081286; 5081287 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A
- Hazard / reason
- increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
- Distribution
- International distribution in the countries of Aland Islands, Australia, Belgium, Denmark, Finland, Germany, Ireland, Italy, Liechtenstein, Netherlands, New Zealand, Norway, Portugal, Spain, Sweden, Switzerland, United Kingdom.
- Code information
- Lot Code: UDI-DI: 04260102136120; 04260102136151
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 4260102136120