Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in…

Title shortened; the full text is below and at the official source.

Agency
U.S. Food and Drug Administration
Recall number
Z-1084-2026
Category
Medical devices
Recall date
2025-12-15
Enforcement report date
2026-01-21
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
Diagnostica Stago, Inc.

From the FDA Enforcement Report dated 2026-01-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Asserachrom HPIA. Model/Catalog Number: 00615. The Asserachrom HPIA kit is intended for the qualitative detection in plasma or serum of the anti-heparin-platelet factor 4 (PF4) antibodies generated during the type II heparin-induced thrombocytopenia (HIT type II) by ELISA.
Hazard / reason
Potential for defect for Asserachrom HPIA lot 271288, which can affect every well in the impacted strip and produce negative results.
Quantity
438 units
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, HI, IA, MA, MI, NJ, NV, NY, OH, TN, WI and the countries of Belgium, Canada, Denmark, Algeria, Spain, Finland, France, Great Britain, Italy, Lebanon, Saudi Arabia, Uganda.
Code information
Reference Number: 00615. UDI: (01)03607450006155(11)240531(17)260531(10)271288(241)00615. Lot Number: 271288.
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands