CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1099-2026
- Category
- Medical devices
- Recall date
- 2025-10-10
- Enforcement report date
- 2026-01-21
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Thoratec LLC
From the FDA Enforcement Report dated 2026-01-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1099-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- CentriMag Blood Pump, REF: 102953, 201-20003, 201-90010, 201-90016; CentriMag RVAD HDE KIT, REF: 104318; CentriMag Pre-connected Pack, REF: CMAEK00; CentriMag VAD KIT, REF: 201-51006, 201-51007, 201-90001 201-90001T; Blood Pump with CentriMag Acute Circulatory Support System for ECMO, REF: CMAEK01. CentriMag Blood Pump is a component to be used with CentriMag Acute Circulatory Support System
- Hazard / reason
- Acute circulatory support system pump may not fit into the motor and reports received of the pump not being securely locked into the motor using the screw-in locking feature. Failure to properly lock the pump into motor my not cause pump not inserted alarm, and could lead to particulate, increased heat, anxiety, inconvenience, delay in hemodynamic/ECMO support, stroke, embolism, death.
- Quantity
- 8399
- Distribution
- Worldwide - US Nationwide distribution including in the states of MN, PA, OK, FL, CA, OR, IL, NY, DE, GA, NJ, NC, WI, OH, AZ, AR, TX, TN, MO, MA, MT, KY, NE, WV, VA, ME, ND, AL, CO, DC, MS, IN, MI, CT, SC, WA, NM, MD, HI, ID, IA, LA, UT, RI, NV and the countries of Algeria, Australia, Austria, Bahrain, CANADA, Cayman Islands, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Hong Kong, Italy, Jordan, Kuwait, Lithuania, Macedonia, Malaysia, Mexico, Netherlands, Oman, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Code information
- UDI-DI: 07640135140627, 07640135140603, 05415067036414 IFU: Part(Name) Revision: For US: 1. ARTEN600254920 (IFU CentriMag Blood Pump ABT) Rev C 2. ARTEN600254958 (IFU PediMag Blood Pump US ABT) Rev C 3. ARTEN100174794 (CentriMag Adult Pre-connected Pack IFU, US) Rev B 4. ARTEN600230157 (IFU CentriMag Blood Pump ECMO US) Rev D FOR OUS: ARTEN600254935 (IFU CentriMag Blood Pump CAN ABT) Rev C ARTEN600254936 (IFU PediVas Blood Pump Canada ABT) Rev C ARTEN600332205 (IFU CMag VAS-ECMO Pump OUS) Rev A ARTEN600332207 (IFU PVAS VAS-ECMO Pump OUS) Rev A REF 201-20003, 104318, 201-51006, 201-51007, 201-90001 201-90001T don't have associated un-expired batches/serial numbers. REF/Batch(Serial): 102953/10007334, 10007335, 10007336, 10012012, 10012013, 10012014, 10012015, 10012016, 10036179, 10036180, 10036181, 10037990, 10037992, 10037993, 10037995, 10037996, 10046028, 10046029, 10046030, 10046031, 10046032, 10046033, 10058754, 10058755, 10058756, 10061673, 10061675, 10061676, 10061677, 10065946, 10065947, 10065954, 10065955, 10065956, 10071358, 10071359, 10071361, 10073780, 10073781, 10073783, 10082053, 10082055, 10103208, 10103209, 10106345, 10106346, 10106347, 10106348, 10120074, 10120075, 10120076, 10120077, 10122157, 10122158, 10122159, 10125562, 10125563, 10125564, 10125565, 10138336, 10138337, 10138338, 10138339, 10145124, 10145125, 10145126, 10145127, 10164040, 10164041, 10164044, 10168193, 10168194, 10168195, 10168196, 10168197, 10168199, 10168200, 10168201, 10168202,…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 7640135140627
- 7640135140603
- 5415067036414