Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG

Agency
U.S. Food and Drug Administration
Recall number
Z-1100-2026
Category
Medical devices
Recall date
2025-12-19
Enforcement report date
2026-01-21
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
ICU Medical Inc.

From the FDA Enforcement Report dated 2026-01-21. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
Pulsator Arterial Blood Sampling Kit, Item Numbers 4044EG-1 4080P-2 4084P-1 4080PE 4080PG
Hazard / reason
Affected lots of syringes may have a crack on the syringe collar, which may lead to delayed or repeated blood draw/fluid delivery. This issue may also lead to blood leakage during sampling of arterial blood or leakage during fluid delivery.
Quantity
64290 total
Distribution
US Nationwide distribution in the states of CA, FL, IL, IN, MA, MI, MN, MO, NE, NJ, NY, OH, OK, TX, WA.
Code information
1. 4044EG-1 UDI-DI 30351688073062 Lots 6141992 6156228 2. 4080P-2 UDI-DI 30351688006176 Lots 6141999 6142000 6147906 3. 4084P-1 UDI-DI 30351688029540 Lots 6147907 (corrected 1/22/26, removed "-LSM" suffix) 4. 4080PE UDI-DI 30351688039150 Lot 6131839 5. 4080PG UDI-DI 30351688081173 Lots 6131838 6142001 6142002
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Identifiers

Barcodes (UPC/GTIN)
  • 30351688073062
  • 30351688006176
  • 30351688029540
  • 30351688039150
  • 30351688081173

Brands