ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)

Agency
U.S. Food and Drug Administration
Recall number
Z-1120-2026
Category
Medical devices
Recall date
2025-12-15
Enforcement report date
2026-01-28
Classification
Class II
Status as reported by FDA
Ongoing
Manufacturer / firm
ICU Medical, Inc.

From the FDA Enforcement Report dated 2026-01-28. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
ICU Medical Plum Solo Precision IV Pump, List Numbers: 400010401 (US), 400011001 (Philippines)
Hazard / reason
Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.
Distribution
Worldwide distribution - US Nationwide and the country of Philippines.
Code information
400010401 (US), Software Versions: 1.0.2; 400011001 (Philippines), Software Versions: 1.0.0
Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class II

Brands