Allwell Angioplasty Pack REF: IS-30-B1/B
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1129-2026
- Category
- Medical devices
- Recall date
- 2025-12-04
- Enforcement report date
- 2026-01-28
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Merit Medical Systems, Inc.
From the FDA Enforcement Report dated 2026-01-28. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1129-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Allwell Angioplasty Pack REF: IS-30-B1/B
- Hazard / reason
- Inflation device handle may detach from the syringe during procedure.
- Quantity
- 328000 units
- Distribution
- Worldwide - Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY, St. Thomas , Puerto Rico and the countries of Albania, Algeria, Angola, Argentina, Australia, Austria, Bahrain, Barbados, Belgium, Brazil, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Iceland, India, Indonesia, Ireland, Israel, Italy, Jamaica, Japan, Kazakhstan, Korea, Republic of Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya, Luxembourg, Madagascar, Mexico, Monaco, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, San Marino, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Trinidad and Tobago, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian, Republic of Viet Nam.
- Code information
- UDI:00884450653388/ Lot: I2563105 I2563106 I2565007 I2598114 I2598131 I2607891 I2612499 I2618005 I2623057 I2632527 I2632528 I2639728 I2661908 I2669772 I2669774 I2827091 I2827094 I2844360 I2844361 I2897482 I2897483 I2897484 I2897485 I2924663 I2972425 I2972426 I2972427 I2972428 I2999054 I3014683 I3030198 I3048363 I3055799 I3080949 I3089230 I3099150 I3110034 I3129711 I3129712 I3161408 I3268359 I3268361 I3268362
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 884450653388