Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM50, iM60, iM70, iM80 Product Description: The iM…
Title shortened; the full text is below and at the official source.
- Agency
- U.S. Food and Drug Administration
- Recall number
- Z-1145-2026
- Category
- Medical devices
- Recall date
- 2025-10-24
- Enforcement report date
- 2026-01-28
- Classification
- Class II
- Status as reported by FDA
- Ongoing
- Manufacturer / firm
- Edan Diagnostics
From the FDA Enforcement Report dated 2026-01-28. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.
Official source
-
FDA does not publish a separate web page for each recall. Search the FDA Enforcement Report for recall number Z-1145-2026.
- Agency data record (JSON)
Agency wording
The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).
- Product
- Brand Name: EDAN Product Name: Patient Monitor Model/Catalog Number: iM50, iM60, iM70, iM80 Product Description: The iM series Patient Monitor including iM50, iM60, iM70 and iM80 can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormalities so that doctors and nurses can respond to the patient's situation as appropriate. Component: No
- Hazard / reason
- As a result of receiving an FDA It Has Come to Our Attention Letter (IHCTOA Letter) regarding potential cybersecurity issues with various of the firm device.
- Quantity
- 14550
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, KY, PR, FL, CT, IL, PA, RI, NC, NY, LA, TX, MI and the country of Mexico.
- Code information
- EDAN Model: iM50 UDI-DI code: 06944413801622 Serial/Lot Numbers: 002327-M16803350001 002327-M16803350002 360069-M17707110001 333109-M16203920004 333109-M16203920005 333136-M16203910001 333136-M16203910002 333136-M16203910003 333136-M16203910004 333136-M16203910006 333136-M16203910007 333136-M16203910008 333136-M16203910009 333136-M16203910010 333136-M16203910005 333109-M16203920003 333109-M16203920002 333109-M16203920001 333109-M16802350004 333109-M16802350005 333109-M16802350006 333109-M16802350007 333109-M16802350008 333136-M16802360001 333136-M16802360002 333136-M16802360003 333136-M16802360004 333109-M16802350003 333136-M16802360005 333109-M16802350002 333109-M16802350001 333136-M16810700002 333136-M16810700003 333136-M16810700004 333136-M16810700005 333136-M16810700001 333136-M16A02590001 333136-M16A02590002 333136-M16A02590005 333136-M16A02590004 333136-M16A02590003 333136-M16A02590006 333136-M16A02590007 333136-M16A02590008 333136-M16A02590009 333136-M16A02590010 333136-M16B08530006 333136-M16B08530011 333136-M16B08530010 333136-M16B08530009 333136-M16B08530008 333136-M16B08530007 333136-M16B08530014 333136-M16B08530005 333136-M16B08530004 333136-M16B08530003 333136-M16B08530002 333136-M16B08530001 333109-M16B08540003 333109-M16B08540002 333109-M16B08540001 333136-M16B08530015 333136-M16B08530012 333136-M16B08530013 261054-M16C02640001 333136-M16B08550009 333136-M16B08550008 333136-M16B08550007 333136-M16B08550006 333136-M16B08550005 333136-M16B08550002…
Shortened. The full text is at the official source.
- Recall type
- Voluntary: Firm initiated
- Product type
- Devices
- Classification (FDA wording)
- Class II
Identifiers
- Barcodes (UPC/GTIN)
- 6944413801622
- 6944413801646
- 6944413801837
- Model numbers
- iM50
- iM80
- iM60