AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm; Material Number (UPN): M00553660;

Agency
U.S. Food and Drug Administration
Recall number
Z-1158-2026
Category
Medical devices
Recall date
2025-12-19
Enforcement report date
2026-02-11
Classification
Class I
Status as reported by FDA
Ongoing
Manufacturer / firm
Boston Scientific Corporation

From the FDA Enforcement Report dated 2026-02-11. The classification and status are as reported by FDA in that report. This page is not updated as the recall progresses. Check the official source.

Official source

Agency wording

The text below is copied unchanged from the FDA record (U.S. Food and Drug Administration, Enforcement Report (openFDA)).

Product
AXIOS Stent and Electrocautery-Enhanced Delivery System - 20mm x 10mm; Material Number (UPN): M00553660;
Hazard / reason
Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.
Quantity
557 units
Distribution
Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Morocco, Netherlands, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, PERU, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, UAE, and United Kingdom.
Code information
Material Number (UPN): M00553680; UDI-DI: 8714729951179; Lot Numbers: 35961006, , 35961007, , 35962524, , 35967379, , 35968520, 35968521, 35968597, 35968598, 35971077, 35971078, 35971079, 35972100, 35982272, 35982273, 35982274, 35982275, 35982276, 35982277, 35982278, 35982279, 35982420, 35982424, 35982425, 35985313, 35985314, 35993619, 35993720, 35993721, 36109761, 36109763, 36109764, 36109765, 36109766, 36111314, 36111315, 36118993, 36118994, 36122066, 36122067, 36122075, 36132801, 36132802, 36132803, 36132804, 36134972, 36144058, 36144059, 36144120, 36144121, 36174475, 36174476, 36178005, 36187588, 36187589, 36187687, 36187688, 36188887, 36188888, 36188889, 36194446, 36194447, 36197239, 36198580, 36217768, 36217769, 36217770, 36227341, 36227342, 36227343, 36228603, 36228604, 36228605, 36235934, 36236151, 36240589, 36240590, 36241715, 36241716, 36241717, 36251363, 36251364, 36253000, 36253008, 36253041, 36260860, 36260861, 36260862, 36261599, 36261780, 36261781, 36261834, 36261835, 36280395, 36280396, 36280397, 36281912, 36281913, 36281914, 36283860, 36292682, 36292683, 36292684, 36292689, 36292690, 36292691, 36294495, 36301621, 36301622, 36301623, 36301624, 36302290, 36302291, 36302292, 36319024, 36319025, 36319026, 36320016, 36320017, 36323789, 36323790, 36324720, 36325159, 36325560, 36325568, 36325569, 36332101, 36332102, 36341577, 36342114, 36342418, 36352008, 36352009, 36355049, 36355050, 36355752, 36359567, 36362263, 36362264, 36362265, 36363150, , 36363151, ,…

Shortened. The full text is at the official source.

Recall type
Voluntary: Firm initiated
Product type
Devices
Classification (FDA wording)
Class I

Identifiers

Barcodes (UPC/GTIN)
  • 8714729951179

Brands